A clinical description of ocular response analyzer measurements

46Citations
Citations of this article
57Readers
Mendeley users who have this article in their library.
Get full text

Abstract

PURPOSE. To examine the interrelationships among the Ocular Response Analyzer (ORA; Reichert Ophthalmic Instruments, Buffalo, NY), Goldmann applanation tonometer (GAT), and corneal geometry measurements in a young, healthy sample. METHODS. Central corneal radius, ORA, GAT, and central corneal thickness (CCT) measurements were taken in 99 subjects (age, 21 ± 2 years) who were free of ocular and systemic disease. RESULTS. The mean ± SD corneal hysteresis (CH) and corneal resistance factor (CRF) were 10.4 ± 1.2 and 10.1 ± 1.5 mm Hg, respectively. The Bland-Altman 95% limits of agreement of ORA Goldmann-correlated IOP (ORAg) and ORA corneal-compensated (ORAcc) IOP with reference to GAT were -4.5 to +6.0 and -4.1 to +6.8 mm Hg, respectively. The full equations used by the ORA to calculate ORAcc and CRF were reconstructed. The statistically significant effect of CCT on GAT became redundant if CRF was included in a multivariate regression analysis. Both CH and CRF were associated with CCT (R 2 = 0.252 and 0.290, respectively). CONCLUSIONS. Sample CH and CRF were consistent with those reported in the literature. ORAg and ORAcc agreed poorly with GAT. CRF appears to be at least a partial description of corneal rigidity. The ocular determinants of CH are unclear. (Invest Ophthalmol Vis Sci. 2011;52:2911-2916.

Cite

CITATION STYLE

APA

Lau, W., & Pye, D. (2011). A clinical description of ocular response analyzer measurements. Investigative Ophthalmology and Visual Science, 52(6), 2911–2916. https://doi.org/10.1167/iovs.10-6763

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free