Analysis of H2O in silicate glass using attenuated total refectance (ATR) micro-FTIR spectroscopy

46Citations
Citations of this article
79Readers
Mendeley users who have this article in their library.
Get full text

Abstract

We present a calibration for attenuated total reflectance (ATR) micro-FTIR for analysis of H2O in hydrous glass. A Ge ATR accessory was used to measure evanescent wave absorption by H2O within hydrous rhyolite and other standards. Absorbance at 3450 cm-1 (representing total H 2O or H2Ot) and 1630 cm-1 (molecular H 2O or H2Om) showed high correlation with measured H2O in the glasses as determined by transmission FTIR spectroscopy and manometry. For rhyolite, wt%H2O = 245(±9)· A3450 - 0.22(±0.03) and wt%H 2Om = 235(±11)· A1630 - 0.20(±0.03) where A3450 and A1630 represent the ATR absorption at the relevant infrared wavelengths. The calibration permits determination of volatiles in singly polished glass samples with spot size down to ∼5 μm (for H2O-rich samples) and detection limits of ∼0.1 wt% H2O. Basaltic, basaltic andesite and dacitic glasses of known H2O concentrations fall along a density-adjusted calibration, indicating that ATR is relatively insensitive to glass composition, at least for calc-alkaline glasses. The following equation allows quantification of H 2O in silicate glasses that range in composition from basalt to rhyolite: wt% H2O = (ω· A3450/ρ) + b where ω= 550 ± 21, b = -0.19 ± 0.03, ρ = density, in g/cm3, and A3450 is the ATR absorbance at 3450 cm -1. The ATR micro-FTIR technique is less sensitive than transmission FTIR, but requires only a singly polished sample for quantitative results, thus minimizing time for sample preparation. Compared with specular reflectance, it is more sensitive and better suited for imaging of H2O variations in heterogeneous samples such as melt inclusions. One drawback is that the technique can damage fragile samples and we therefore recommend mounting of unknowns in epoxy prior to polishing. Our calibration should hold for any Ge ATR crystals with the same incident angle (31°). Use of a different crystal type or geometry would require measurement of several H2O-bearing standards to provide a crystal-specific calibration.

Cite

CITATION STYLE

APA

Lowenstern, J. B., & Pitcher, B. W. (2013). Analysis of H2O in silicate glass using attenuated total refectance (ATR) micro-FTIR spectroscopy. American Mineralogist, 98(10), 1660–1668. https://doi.org/10.2138/am.2013.4466

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free