Atmospheric CO 2 monitoring with single-cell NDIR-based analyzers

43Citations
Citations of this article
56Readers
Mendeley users who have this article in their library.

Abstract

We describe CO 2 concentration measurement systems based on relatively inexpensive single-cell non-dispersive infrared CO 2 sensors. The systems utilize signal averaging to obtain precision (1-f in 100 s) of 0.1 parts per million dry air mole fraction (ppm), frequent calibrations and sample drying in order to achieve state-of-the-art compatibility, and can run autonomously for months at a time. Laboratory tests indicate compatibility among four to six systems to be ±0.1 ppm (1-f), and field measurements of known reference-gases yield median errors of 0.01 to 0.17 ppm with 1-f variance of ±0.1 to 0.2 ppm. From May to August 2007, a system co-located with a NOAA-ESRL dual-cell NDIR system at the WLEF tall tower in Wisconsin measured daytime-only daily averages of CO 2 that differ by 0.26 ± 0.15 ppm (median ± 1 f), and from August 2005 to April 2011 a system co-located with weekly NOAA-ESRL network flask collection at Niwot Ridge, Colorado measured coincident CO 2 concentrations that differed by g -0.06 ± 0.30 ppm (in/i Combining double low line 585). Data from these systems are now supporting a wide range of analyses and this approach may be applicable in future studies where accuracy and initial cost of the sensors are priorities. © Author(s) 2011. CC Attribution 3.0 License.

Cite

CITATION STYLE

APA

Stephens, B. B., Miles, N. L., Richardson, S. J., Watt, A. S., & Davis, K. J. (2011). Atmospheric CO 2 monitoring with single-cell NDIR-based analyzers. Atmospheric Measurement Techniques, 4(12), 2737–2748. https://doi.org/10.5194/amt-4-2737-2011

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free