Blood flow rate and access recirculation in hemodialysis

22Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.
Get full text

Abstract

We studied the effect of extracorporeal blood flow rate (BFR) on access recirculation (recirc) in 19 hemodialysis patients. BUN was determined in simultaneous peripheral (P), arterial (A), and venous (V) blood obtained at BFRs of 200, 400 and 600 ml/min. Percent recirc was calculated for each BFR using the formula (P-A)/(P-V) x 100. Venous drip-chamber (VP) and pre-blood-pump (AP) pressures were measured at each BFR. Fistulograms were performed in 10 patients, and stenoses were identified in 5, all at the proximal (arterial) end of the access. Recirc increased with increasing BPR from 200 to 400 ml/min but increased little from 400 to 600 ml/min. At all BFRs recirc in the stenotic patients was higher than that of non-stenotic or unstudied patients. Urea clearance, corrected for recirc, rose with blood flow both in stenotic and non-stenotic patients. There were no differences in AP or in VP between stenotic and non-stenotic patients. At BFR ≥ 400 ml/min, a recirc threshold of 15% identified stenoses with sensitivity 100% and specificity 71%. We conclude (1) recirc increases with increasing BFR but not enough to outweight the concomitant increase in urea clearance; (2) significant access stenosis and recirc may be present even with low VP; (3) recirc was associated with arterial side stenoses; (4) at BFR ≥ 400 ml/min, access stenosis is associated with recirc < 15%.

Cite

CITATION STYLE

APA

Daniels, I. D., Berlyne, G. M., & Barth, R. H. (1992). Blood flow rate and access recirculation in hemodialysis. International Journal of Artificial Organs, 15(8), 470–474. https://doi.org/10.1177/039139889201500805

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free