The Characterization of an Electrothermal Vaporization-Direct Current Plasma Atomic Emission Spectrometer for the Determination of Boron, Cadmium, Copper, Iron, and Lead

0Citations
Citations of this article
1Readers
Mendeley users who have this article in their library.
Get full text

Abstract

A graphite rod electrothermal vaporizer used to introduce microliter-sized samples into a direct current plasma (DCP) atomic emission spectrometer is reported in this work. Several important experimental conditions were found to be important in achieving good analytical performance from the electrothermal vaporization (ETV-DCP) system. A combination of lowered plasma electrode sleeve gas flow rates, when compared to that commonly used in the DC plasma jet, and relatively low carrier gas flow rates resulted in very good analytical performance. Relative precision for the ETV-DCP instrument using B, Cd, Cu, Fe, and Pb solutions ranged between 4% and 10%. Limits of detection (LOD) lower than 100pg were achieved for these elements, roughly an order of magnitude better than other ETV-DCP studies that used commercial graphite boat/furnace combinations. In addition, good calibration linearity was observed, with 2-4 orders of magnitude linearity for the elements investigated.

Cite

CITATION STYLE

APA

Schroeder, T. S., Mahon, A. R., Conver, T. S., Hahn, T., Ramsay, K., Ambrose, T., … McCurdy, D. L. (2004). The Characterization of an Electrothermal Vaporization-Direct Current Plasma Atomic Emission Spectrometer for the Determination of Boron, Cadmium, Copper, Iron, and Lead. Spectroscopy Letters, 37(2), 175–190. https://doi.org/10.1081/SL-120030852

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free