Development and validation of a rapid HPLC method for the determination of ketotifen in pharmaceuticals

18Citations
Citations of this article
23Readers
Mendeley users who have this article in their library.
Get full text

Abstract

In the present study, a simple, sensitive, rapid, and stability-indicating high performance liquid chromatographic (HPLC) method with ultraviolet detection for the analysis of ketotifen was developed and validated. The method was applied to the determination of ketotifen in pharmaceutical formulations (tablets and syrups). The HPLC method utilized isocratic elution technique with a reversed phase C8 column, detection at 297 nm and a mixture of methanol, triethylamine phosphate buffer (pH 2.8; 0.04 M), and tetrahydrofuran (43: 55: 2, v/ v/v) as mobile phase at a flow rate of 1.2 mL/min. Total analysis time was about 7 min with typical retention time of ketotifen of about 5 min. The method was validated for selectivity, linearity, accuracy, and precision following International Conference of Harmonization, 1996 (ICH) recommendations. Due to its simplicity and accuracy, the method can be used for routine quality control analysis. Copyright © Taylor & Francis Group, LLC.

Cite

CITATION STYLE

APA

Elsayed, M. M. A. (2006). Development and validation of a rapid HPLC method for the determination of ketotifen in pharmaceuticals. Drug Development and Industrial Pharmacy, 32(4), 457–461. https://doi.org/10.1080/03639040500529135