RapidArc ™ treatment verification using polymer gel dosimetry
Measurement of particle size in polymer gel dosimeters
using spectrophotometry
Heather M. Whitney and John C. Gore
Institute of Imaging Science, Vanderbilt University, Nashville, TN, USA
1. Purpose
Polymer gel dosimeters have proven to be useful for three-dimensional radiation dosimetry in
clinical radiation therapy. However, more study and development work is needed to quantify and
further understand the dose response mechanism, with the goal of improving the dosimeters. In
this work we seek to determine the diameter of the polymer particles formed upon irradiation, for
a range of irradiation dose rates. It is expected that this information will be used to further
understand the actual formation of polymer in the gelatin matrix, and how it affects the relaxation
of water in the dosimeter as measured with magnetic resonance imaging or optical methods.
2. Materials and Methods
MAGIC-2 gel dosimeters were made using standard preparation techniques [1]. The specific
formulation is comprised of 87% water, 4% methacrylic acid, 9% gelatin, [Cu2] = 1.738 x 10-5 M,
and a ratio of ascorbic acid to [Cu2+] of 1000:1. After preparation, the gels were poured into
plastic cuvettes and sealed with plastic toppers. The boundaries of the cuvettes and toppers were
dipped in sealing wax to prevent leakage of oxygen into the gel. The gels were stored in a closed
styrofoam container, refrigerated overnight, and irradiated the next day using a Therapax
orthovoltage x-ray unit. Samples were refrigerated for three weeks and then analyzed with a UV-
VIS spectrophotometer. Absorbance versus wavelength was measured for a wavelength range of
290 through 900 nm.
Results were analyzed using a method previously implemented with BANG gels [2].
Assuming that the polymer particles formed in the MAGIC gel are spherical, that their index of
refraction is 1.5, and light absorption is negligible, the particle diameter can be determined by the
relation
S
O
n
ka
a 0max2 (1)
where a is the maximum particle radius, k is the wave number in the medium, kamax-is the value of
ka at which Qsca, the Mie-Debye efficiency factor, is maximized, Ȝ0-is the wavelength at which
maximum turbidity is observed, and n is the index of refraction. From the literature, kamax is
considered to be 4.34 [3]. Particle sizes were determined for the MAGIC gel dosimeters as a
function of dose, over a range of dose rates.
Institute of Physics Publishing Journal of Physics: Conference Series 56 (2006) 160–163
doi:10.1088/1742-6596/56/1/017 4th International Conference on Radiotherapy Gel Dosimetry
160© 2006 IOP Publishing Ltd
using spectrophotometry
Heather M. Whitney and John C. Gore
Institute of Imaging Science, Vanderbilt University, Nashville, TN, USA
1. Purpose
Polymer gel dosimeters have proven to be useful for three-dimensional radiation dosimetry in
clinical radiation therapy. However, more study and development work is needed to quantify and
further understand the dose response mechanism, with the goal of improving the dosimeters. In
this work we seek to determine the diameter of the polymer particles formed upon irradiation, for
a range of irradiation dose rates. It is expected that this information will be used to further
understand the actual formation of polymer in the gelatin matrix, and how it affects the relaxation
of water in the dosimeter as measured with magnetic resonance imaging or optical methods.
2. Materials and Methods
MAGIC-2 gel dosimeters were made using standard preparation techniques [1]. The specific
formulation is comprised of 87% water, 4% methacrylic acid, 9% gelatin, [Cu2] = 1.738 x 10-5 M,
and a ratio of ascorbic acid to [Cu2+] of 1000:1. After preparation, the gels were poured into
plastic cuvettes and sealed with plastic toppers. The boundaries of the cuvettes and toppers were
dipped in sealing wax to prevent leakage of oxygen into the gel. The gels were stored in a closed
styrofoam container, refrigerated overnight, and irradiated the next day using a Therapax
orthovoltage x-ray unit. Samples were refrigerated for three weeks and then analyzed with a UV-
VIS spectrophotometer. Absorbance versus wavelength was measured for a wavelength range of
290 through 900 nm.
Results were analyzed using a method previously implemented with BANG gels [2].
Assuming that the polymer particles formed in the MAGIC gel are spherical, that their index of
refraction is 1.5, and light absorption is negligible, the particle diameter can be determined by the
relation
S
O
n
ka
a 0max2 (1)
where a is the maximum particle radius, k is the wave number in the medium, kamax-is the value of
ka at which Qsca, the Mie-Debye efficiency factor, is maximized, Ȝ0-is the wavelength at which
maximum turbidity is observed, and n is the index of refraction. From the literature, kamax is
considered to be 4.34 [3]. Particle sizes were determined for the MAGIC gel dosimeters as a
function of dose, over a range of dose rates.
Institute of Physics Publishing Journal of Physics: Conference Series 56 (2006) 160–163
doi:10.1088/1742-6596/56/1/017 4th International Conference on Radiotherapy Gel Dosimetry
160© 2006 IOP Publishing Ltd
Page 3
Figure 1. Absorbance spectra for four dosimeters and a single dose rate.
Figure 2. Absorbance versus dose for a single dosimeter.
161
Figure 2. Absorbance versus dose for a single dosimeter.
161
Page 4
3. Results
The results for a sample spectrophotometric experiment are shown in Figure. Absorbance
spectra are given for four different dose levels. The wavelength of maximum turbidity is chosen
from these data. Figure 2 displays the measured absorbance versus dose for a given dosimeter,
measured at three different wavelengths of light. In each case, absorbance increases with dose.
Figure 3 shows the results for maximum particle diameter versus dose for a range of dose rates
from 0.22 to 2.60 Gy/min. Particle size varied from 270 to 334 nm over a dose range of 0 to 40
Gy. There is a clear increase of particle size with dose, but the correspondence of this behaviour
to dose rate is not yet clear.
4. Conclusions
These experiments show an increase in the diameter of polymer formed in the MAGIC-2 gel
dosimeter associated with increased applied dose. Further studies will compare these data to
molecular weight measurements.
5. References
[1] Fong P M, Keil D C, Does M D, Gore J C 2001 Polymer gels for magnetic resonance
imaging of radiation dose distributions at normal room atmosphere Phys. Med. Biol. 46
3105-3113
[2] Maryanski M J, Zastavker Y Z and Gore J C 1996 Radiation dose distributions in three
dimensions from tomographic optical density scanning of polymer gels. 2. Optical
properties of the BANG polymer gel Phys. Med. Biol. 41 2705-2717
[3] Wickramasinghe N C 1973 Light Scattering Functions for Small Particles New York: John
Wiley & Son, Inc
162
The results for a sample spectrophotometric experiment are shown in Figure. Absorbance
spectra are given for four different dose levels. The wavelength of maximum turbidity is chosen
from these data. Figure 2 displays the measured absorbance versus dose for a given dosimeter,
measured at three different wavelengths of light. In each case, absorbance increases with dose.
Figure 3 shows the results for maximum particle diameter versus dose for a range of dose rates
from 0.22 to 2.60 Gy/min. Particle size varied from 270 to 334 nm over a dose range of 0 to 40
Gy. There is a clear increase of particle size with dose, but the correspondence of this behaviour
to dose rate is not yet clear.
4. Conclusions
These experiments show an increase in the diameter of polymer formed in the MAGIC-2 gel
dosimeter associated with increased applied dose. Further studies will compare these data to
molecular weight measurements.
5. References
[1] Fong P M, Keil D C, Does M D, Gore J C 2001 Polymer gels for magnetic resonance
imaging of radiation dose distributions at normal room atmosphere Phys. Med. Biol. 46
3105-3113
[2] Maryanski M J, Zastavker Y Z and Gore J C 1996 Radiation dose distributions in three
dimensions from tomographic optical density scanning of polymer gels. 2. Optical
properties of the BANG polymer gel Phys. Med. Biol. 41 2705-2717
[3] Wickramasinghe N C 1973 Light Scattering Functions for Small Particles New York: John
Wiley & Son, Inc
162
Page 5
Figure 3. Calculated maximum particle diameter.
163
163
Sign up today - FREE
Mendeley saves you time finding and organizing research. Learn more
- All your research in one place
- Add and import papers easily
- Access it anywhere, anytime
Start using Mendeley in seconds!
Readership Statistics
2 Readers on Mendeley
by Discipline
100% Physics
by Academic Status
50% Ph.D. Student
50% Assistant Professor
by Country
50% Spain
50% United States