A Comparison of In-Person Examination and Video Otoscope Imaging for Tympanostomy Tube Follow-Up

58Citations
Citations of this article
43Readers
Mendeley users who have this article in their library.
Get full text

Abstract

The objective of this study was to determine if video otoscope still images (640 × 480 pixel resolution) of the tympanic membrane following surgical placement of tympanostomy tubes are comparable to an in-person microscopic examination. Forty patients having undergone tympanostomy tube placement in both ears were independently examined in-person by two otolaryngologists and imaged using a video otoscope and telemedicine software package. The two physicians later reviewed images at 6 and 12 weeks. Physical examination findings and diagnosis were documented and compared for their concordance using kappa statistics. For both physicians, the intraprovider concordance between the in-person examination and the corresponding image review was high for each of the physical examination findings: Tube In 93-94% (K 0.85-0.87), Tube Patent 86-93% (K 0.74-0.85), Drainage 94-98% (K 0.42-0.66), Perforation 85-98% (K 0.40-0.84), Granulation 95-99% (K -0.01 to 0.00), Middle Ear Fluid 89-91% (K -0.03 to 0.50), and Retracted 89-94% (K 0.13-0.43). These agreement rates are similar to the normal interprovider concordance observed when two physicians independently examined the same patient in-person for physical exam findings: Tube In 96% (K 0.93), Tube Patent 94% (K 0.88), Drainage 96% (K 0.56), Perforation 90% (K 0.60), Granulation 96% (K 0.39), Middle Ear Fluid 88% (K 0.14), and Retracted 91% (K 0.43). For both physicians, the intraprovider diagnostic concordance between the in-person examination and the corresponding image review was high 79-85% (K 0.67-0.76). The interprovider diagnostic concordance for the in-person exam was 88% (K 0.81). The interprovider diagnostic concordance when two physicians independently reviewed all images was 84% (K 0.74), and 89% (K 0.80) when poor images were excluded. This study demonstrates that physician review of video otoscope images is comparable to an in-person microscopic examination. Store-and-forward video otoscopy may be an acceptable method of following patients post-tympanostomy tube placement.

Cite

CITATION STYLE

APA

Patricoski, C., Kokesh, J., Ferguson, A. S., Koller, K., Zwack, G., Provost, E., & Holck, P. (2003). A Comparison of In-Person Examination and Video Otoscope Imaging for Tympanostomy Tube Follow-Up. Telemedicine and E-Health, 9(4), 331–344. https://doi.org/10.1089/153056203772744653

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free