Accuracy of a novel real-time continuous glucose monitoring system: a prospective self-controlled study in thirty hospitalized patients with type 2 diabetes

1Citations
Citations of this article
8Readers
Mendeley users who have this article in their library.

Abstract

Aims: The present study aimed to investigate the accuracy of the Glunovo® real-time continuous glucose monitoring system (rtCGMS). Methods: We conducted a 14-day interstitial glucose level monitoring using Glunovo® rtCGMS on thirty hospitalized patients with type 2 diabetes. The flash glucose monitoring (FGM) was used as a self-control. Consistency tests, error grid analysis, and calculation of the mean absolute relative difference (MARD) were performed using R software to assess the accuracy of Glunovo® rtCGMS. Results: Glunovo® exhibited an overall MARD value of 8.89% during hospitalization, compared to 10.42% for FGM. The overall percentages of glucose values within ±10%/10, ± 15%/15, ± 20%/20, ± 30%/30, and ±40%/40 of the venous blood glucose reference value were 63.34%, 81.31%, 90.50%, 97.29%, and 99.36% for Glunovo®, respectively, compared with 61.58%, 79.63%, 88.31%, 96.22% and 99.23% for FGM. The Clarke Error Grid Analysis showed that 99.61% of Glunovo® glucose pairs and 100.00% of FGM glucose pairs within zones A and B. Conclusion: Our study confirms the superior accuracy of Glunovo® in monitoring blood glucose levels among hospitalized patients with type 2 diabetes.

Cite

CITATION STYLE

APA

Ge, S., Zhang, H., Wang, J., Li, H., Su, X., Ding, D., & Ma, J. (2024). Accuracy of a novel real-time continuous glucose monitoring system: a prospective self-controlled study in thirty hospitalized patients with type 2 diabetes. Frontiers in Endocrinology, 15. https://doi.org/10.3389/fendo.2024.1374496

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free