Abstract
A rapid, simple and validated reversed-phase high-performance liquid chromatographic method has been developed for analysis of aripiprazole in tablet dosage form. Aripiprazole was separated on an ODS analytical column with a 40:60 (v/v) mixture of acetonitrile and triethanolamine buffer (5 mM, pH 3.5 ± 0.05 adjusted by addition of 85% phosphoric acid) as mobile phase at a flow rate of 1.5 mL -1. The effluent was monitored by UV detection at 254 nm. Calibration plots were linear in the range of 20 to 60 μg mL-1 and the LOD and LOQ were 0.411 and 1.248 μg mL-1, respectively. The high recovery and low relative standard deviation confirm the suitability of the method for routine quality control determination of aripiprazole in tablets.
Author supplied keywords
Cite
CITATION STYLE
Kalaichelvi, R., Thangabalan, B., & Srinivasa Rao, D. (2010). Validated RP-HPLC method for analysis of aripiprazole in a formulation. E-Journal of Chemistry, 7(3), 827–832. https://doi.org/10.1155/2010/935279