A failure of M-Entropy™ to correctly detect burst suppression leading to sevoflurane overdosage

23Citations
Citations of this article
27Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Electroencephalogram depth of anaesthesia monitors are increasingly being used, with the aim of reducing awareness during anaesthesia. Most literature concentrates on the ability of these monitors to predict when a patient is likely to be aware. This case report highlights the opposite problem, where the monitor (M-Entropy™) indicated an awake state but the patient was in fact deeply anaesthetised. If the anaesthetist is unable to interpret the raw electroencephalogram and understand the limitations of the monitor being used, excessive doses of anaesthetic may be given with potentially serious consequences.

Cite

CITATION STYLE

APA

Hart, S. M., Buchannan, C. R., & Sleigh, J. W. (2009). A failure of M-EntropyTM to correctly detect burst suppression leading to sevoflurane overdosage. Anaesthesia and Intensive Care, 37(6), 1002–1004. https://doi.org/10.1177/0310057x0903700619

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free