Multislice perfusion and perfusion territory imaging in humans with separate label and image coils

119Citations
Citations of this article
80Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

An arterial spin labeling technique using separate RF labeling and imaging coils was used to obtain multislice perfusion images of the human brain at 3 T. Continuous RF irradiation at a peak power of 0.3 W was applied to the carotid arteries to adiabatically invert spins. Labeling was achieved without producing magnetization transfer effects since the B1 field of the labeling coil did not extend into the imaging region or couple significant power into the imaging coil. Eliminating magnetization transfer allowed the acquisition of multislice perfusion images of arbitrary orientation. Combining surface coil labeling with a reduced RF duty cycle permitted significantly lower SAR than single coil approaches. The technique was also found to allow selective labeling of blood in either carotid, providing an assessment of the artery's perfusion territory. In normal subjects, these territories were well-defined and localized to the ipsilateral hemisphere.

Cite

CITATION STYLE

APA

Zaharchuk, G., Ledden, P. J., Kwong, K. K., Reese, T. G., Rosen, B. R., & Wald, L. L. (1999). Multislice perfusion and perfusion territory imaging in humans with separate label and image coils. Magnetic Resonance in Medicine, 41(6), 1093–1098. https://doi.org/10.1002/(SICI)1522-2594(199906)41:6<1093::AID-MRM4>3.0.CO;2-0

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free