Evaluation of Low-kVp Low-Volume Iodinated Contrast Protocol for Coronary CT Angiography Using Retrospective ECG Gating

0Citations
Citations of this article
20Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Objective The aim of this study was to evaluate the use of low peak kilovoltage (kVp) low-volume iodinated contrast protocol for performing coronary computed tomography (CT) angiography (CCTA) in patients using retrospective electrocardiogram (ECG) gating. Materials and Methods Hundred prospective patients undergoing CCTA were studied in two groups, A and B, using 70 kilovoltage (kV) and 120 kV protocols with half and standard intravenous volumes of injected iodinated contrast, respectively. All patients had heart rates less than 100 beats/min and body mass index (BMI) less than 31 kg/m 2. Both the groups were evaluated for signal-to-noise (S/N) and contrast-to-noise (C/N) ratios along with radiation dose delivered in millisievert (mSv), and for image quality (IQ), on per patient and per segment basis. Results Patients with group A showed statistically reduced radiation dose of 1.86 mSv compared with 6.86 mSv in group B patients. Marked reduction in image noise with statistically improved S/N and C/N ratios in all coronary vessels was seen in group A. S/N ratios in group A were 20.25, 18.68, 19.04, 17.41, and 18.69 for aorta, left main, left anterior descending, right coronary, and left circumflex arteries while they were 13.34, 11.12, 10.96, 9.74, and 8.67 in group B patients. C/N ratios were also higher in all vessels in group A patients, that is, 19.48, 19.48, 19.04, 19.48, and 17.68, compared with group B patients, who had 12.43, 10.03, 9.23, 9.57, and 8.23 ratios (p < 0.0001). No significant difference in IQ per patient and per vessel was seen between both the groups. Discussion Retrospective ECG-gated low-kVp low-volume iodinated contrast protocol provides good diagnostic quality angiograms in patients with BMI up to 31 kg/m 2 and with heart rates of less than 100 beats/min with three times reduced radiation dose. The reduced volume of contrast reduces the cost as well as the chance of contrast-induced nephropathy.

Cite

CITATION STYLE

APA

Kapoor, A., Mahajan, G., & Kapoor, A. (2021). Evaluation of Low-kVp Low-Volume Iodinated Contrast Protocol for Coronary CT Angiography Using Retrospective ECG Gating. Indian Journal of Radiology and Imaging, 31(4), 910–916. https://doi.org/10.1055/s-0041-1741102

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free