2D EIT for biomedical imaging: Design, measurement, simulation, and image reconstruction

2Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.
Get full text

Abstract

A 2D electrical impedance tomography (EIT) system has been developed at Duke University as an experimental system to test the forward and inverse algorithms for EIT application. The forward model is based on the 2nd-order finite element method (FEM), while the image reconstruction is based on the distorted Born iterative method (DBIM). The major contributions of this work are the application of the higherorder FEM as a forward solver, and the DBIM as an inverse solver to the integrated EIT system. The forward model has been validated with the measured data to within 0.5% accuracy. Excellent images have been reconstructed with these collected EIT data sets. © 2007 Wiley Periodicals, Inc.

Cite

CITATION STYLE

APA

Lim, K. H., Shi, G., McCarter, K., George, R., Ybarra, G., Joines, W. T., … Liu, Q. H. (2007). 2D EIT for biomedical imaging: Design, measurement, simulation, and image reconstruction. Microwave and Optical Technology Letters, 49(12), 2989–2998. https://doi.org/10.1002/mop.22938