Study of flow rate induced measurement error in flow-through nano-hole plasmonic sensor

2Citations
Citations of this article
15Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Flow-through gold film perforated with periodically arrayed sub-wavelength nano-holes can cause extraordinary optical transmission (EOT), which has recently emerged as a label-free surface plasmon resonance sensor in biochemical detection by measuring the transmission spectral shift. This paper describes a systematic study of the effect of microfluidic field on the spectrum of EOT associated with the porous gold film. To detect biochemical molecules, the sub-micron-thick film is free-standing in a microfluidic field and thus subject to hydrodynamic deformation. The film deformation alone may cause spectral shift as measurement error, which is coupled with the spectral shift as real signal associated with the molecules. However, this microfluid-induced measurement error has long been overlooked in the field and needs to be identified in order to improve the measurement accuracy. Therefore, we have conducted simulation and analytic analysis to investigate how the microfluidic flow rate affects the EOT spectrum and verified the effect through experiment with a sandwiched device combining Au/Cr/Si3N4 nano-hole film and polydimethylsiloxane microchannels. We found significant spectral blue shift associated with even small flow rates, for example, 12.60 nm for 4.2 μl/min. This measurement error corresponds to 90 times the optical resolution of the current state-of-the-art commercially available spectrometer or 8400 times the limit of detection. This really severe measurement error suggests that we should pay attention to the microfluidic parameter setting for EOT-based flow-through nano-hole sensors and adopt right scheme to improve the measurement accuracy.

Cite

CITATION STYLE

APA

Tu, L., Huang, L., Wang, T., & Wang, W. (2015). Study of flow rate induced measurement error in flow-through nano-hole plasmonic sensor. Biomicrofluidics, 9(6). https://doi.org/10.1063/1.4936863

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free