Development of a new immunosensor for the detection of dopamine

0Citations
Citations of this article
12Readers
Mendeley users who have this article in their library.

Abstract

Graphite immunoelectrodes as immunosensors using indirect immobilization of a hapten were investigated for their applicability to detect dopamine hydrochloride at low levels. Conditions were optimized to achieve the highest sensitivity using the indirect immobilization of dopamine hydrochloride through a polymerized glutaraldehyde network on microtiter plates using ELISA technique. The conditions were later transferred to the graphite rods (0 0.8 m x 20 mm) and a comparison between the two different sensitivities (IC50 midpoint of test) was carried out. Graphite electrodes showed higher sensitivity towards dopamine than ELISA, since they were able to detect dopamine with a midpoint of test of 0.2 mmol/l while using ELISA they were able to detect dopamine hydrochloride at 2 mmol/l. © 2007 Verlag der Zeitschrift für Naturforschung.

Cite

CITATION STYLE

APA

Fouad, D. (2007). Development of a new immunosensor for the detection of dopamine. Zeitschrift Fur Naturforschung - Section C Journal of Biosciences, 62(7–8), 613–618. https://doi.org/10.1515/znc-2007-7-826