Abstract
A liquid chromatographic method is described for analysis of benzocaine (BZ), a proposed fish anesthetic, in rainbow trout plasma. Mean recoveries of BZ from plasma samples fortified at 44-10 100 ng/mL were 96-100%. The method detection limit is 10 ng/mL, and the limit of quantitation is 37 ng/mL. Acetylation of BZ occurs in whole blood after storage at room temperature (i.e., 21°C) for 10 min. However, no acetylation of BZ was detected in plasma samples held at room temperature for 4 h. Mean method precision for plasma samples with incurred BZ residue is similar to that for fortified samples in the same concentration range (relative standard deviations of 0.9 and 1.2%, respectively).
Cite
CITATION STYLE
Bernardy, J. A., Coleman, K. S., Stehly, G. R., & Gingerich, W. H. (1996). Determination of Benzocaine in Rainbow Trout Plasma. Journal of AOAC International, 79(3), 623–635. https://doi.org/10.1093/jaoac/79.3.623