During prostate brachytherapy, C-arm flouroscopy images are used for a qualitative assessment of the procedure. Three dimensional reconstruction of the implanted seeds can be used for intraoperative dosimetry and quantitative assessment. Accurate C-arm pose estimation is necessary for 3D localization of the seeds. We propose to measure the C-arm rotation angles and computationally compensate the inevitable C-arm translational motion to estimate the pose. We compensate the oscillation, sagging and wheel motion of the C-arm using a three-level optimization algorithm, without which the reconstruction can fail. We validated our approach on simulated and 10 data sets from 5 patients and gained on average 99.1% success rate, 0.33 mm projection error and computation time of less than one minute per patient, which are clinically excellent results. © 2010 Springer-Verlag.
CITATION STYLE
Dehghan, E., Lee, J., Moradi, M., Wen, X., Fichtinger, G., & Salcudean, S. E. (2010). Prostate brachytherapy seed reconstruction using C-arm rotation measurement and motion compensation. In Lecture Notes in Computer Science (including subseries Lecture Notes in Artificial Intelligence and Lecture Notes in Bioinformatics) (Vol. 6361 LNCS, pp. 283–290). https://doi.org/10.1007/978-3-642-15705-9_35
Mendeley helps you to discover research relevant for your work.