The effect of single use laryngoscopy equipment on illumination for tracheal intubation

14Citations
Citations of this article
8Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

We measured the illumination generated by all 30 Macintosh size 3 laryngoscopes in our department with a lux-meter and a standardised laryngoscope holding tube. We found a large range in illumination generated (65-3130 lx). We then measured the effect on the illumination for each laryngoscope by covering the blade with a cover (LaryGard). In every case, the illumination was reduced by the LaryGard, the mean (SD) reduction was 67 (19)%. When we compared the illumination generated by a disposable laryngoscope blade with the same power source, we found that the illumination was reduced less than with the standard Macintosh covered by a LaryGard.

Cite

CITATION STYLE

APA

Anderson, K. J., & Bhandal, N. (2002). The effect of single use laryngoscopy equipment on illumination for tracheal intubation. Anaesthesia, 57(8), 773–777. https://doi.org/10.1046/j.1365-2044.2002.02691.x