Abstract
We measured the illumination generated by all 30 Macintosh size 3 laryngoscopes in our department with a lux-meter and a standardised laryngoscope holding tube. We found a large range in illumination generated (65-3130 lx). We then measured the effect on the illumination for each laryngoscope by covering the blade with a cover (LaryGard). In every case, the illumination was reduced by the LaryGard, the mean (SD) reduction was 67 (19)%. When we compared the illumination generated by a disposable laryngoscope blade with the same power source, we found that the illumination was reduced less than with the standard Macintosh covered by a LaryGard.
Author supplied keywords
Cite
CITATION STYLE
Anderson, K. J., & Bhandal, N. (2002). The effect of single use laryngoscopy equipment on illumination for tracheal intubation. Anaesthesia, 57(8), 773–777. https://doi.org/10.1046/j.1365-2044.2002.02691.x