Validity, intra- and inter-observer reliability of automated devices for the assessment of ankle brachial index using photo-plethysmography

10Citations
Citations of this article
24Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background: Ankle-brachial-Index (ABI) measured by manual Dopplersonography is an easily assessable marker of global cardiovascular risk. The aim of this study was to establish novel photo-plethysmography (PPG)-based ABI assessments in an epidemiologic context and to compare its results with those of Doppler. Methods: Two devices for PPG-based ABI assessments (Vicorder, Vascular Explorer) were tested and compared against Doppler in 56 putatively healthy subjects. We determined acceptance, time requirements, agreement of repeat measurements, agreement with Doppler and intra- and inter-observer concordances for both devices and compared the results. Differences between cuff inflation- and deflation-based methods were also studied for Vascular Explorer. Results: Acceptance was similar for both devices but Vascular Explorer was more time consuming. Agreement of multiple measurements was moderate for both methods highlighting the importance of measurement replicates. Both automated devices showed significantly higher ABI compared to Doppler which can be traced back to higher brachial pressures (Vicorder) or higher ankle pressures (Vascular Explorer). This effect is more pronounced for Vascular Explorer but can be ameliorated using the deflation method of measurement. Intra-observer concordances were similar. Inter-observer concordance was non-significantly better for Vicorder. Conclusions: Both devices proved to be feasible in epidemiologic studies, but compared to Doppler, do not constitute an advantage regarding time requirement and accuracy of ABI assessment. Since PPG-based ABI values are inflated compared to Doppler, it will be necessary to adjust Doppler-based cut-offs for risk stratification. © 2013 Teren et al.; licensee BioMed Central Ltd.

Cite

CITATION STYLE

APA

Teren, A., Beutner, F., Wirkner, K., Loeffler, M., & Scholz, M. (2013). Validity, intra- and inter-observer reliability of automated devices for the assessment of ankle brachial index using photo-plethysmography. BMC Cardiovascular Disorders, 13. https://doi.org/10.1186/1471-2261-13-81

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free