Abstract
In this study, we demonstrated that the automated hematology analyzer, XN-31, can detect red blood cells infected with Babesia caballi and Theileria equi in about 1 minute. We evaluated the diagnostic performance of the XN-31 analyzer for equine piroplasmosis, providing evidence of its potential as a diagnostic tool for this disease.
Cite
CITATION STYLE
APA
Ochi, A., Toya, Y., Sengoku, M., Tsuchiya, S., Kishi, D., & Ueno, T. (2024). In vitro evaluation of the automated hematology analyzer XN-31 for rapid diagnosis of equine piroplasmosis. Microbiology Spectrum, 12(10). https://doi.org/10.1128/spectrum.00582-24