A new gamma camera with a Gas Electron Multiplier

1Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

Abstract

We have developed a prototype gaseous gamma camera with a Gas Electron Multiplier (GEM) for biomolecular analysis and medical applications. The system consists of three devices: a detector, an Ethernet hub, and a PC. The detector consists of a GEM-chamber and integrated readout electronics. The GEM-chamber consists of three parts: the gamma-ray conversion and amplification layers consisting of GEM foils, and a readout component. To increase gamma-ray-detection efficiency, we constructed a new type of GEM foil. The gamma-ray conversion layers are plated with gold 3 μm thick on both surfaces. In order to test the performance of the prototype camera, several measurements were performed with a phantom filled with 99mTc (141 keV) as the radioactive source, and two-dimensional (2D) images were obtained using a pinhole collimator. The camera showed good capability to resolve objects of a few mm2. We believe that this system will offer new knowledge in biomolecular analysis and medical applications. © 2012 IOP Publishing Ltd and SISSA.

Cite

CITATION STYLE

APA

Koike, T., Uno, S., Uchida, T., Sekimoto, M., Murakami, T., Shoji, M., … Nakano, E. (2012). A new gamma camera with a Gas Electron Multiplier. Journal of Instrumentation, 7(1). https://doi.org/10.1088/1748-0221/7/01/C01078