A 74-year-old female presented with intraventricular and subarachnoid hemorrhage due to probable rupture of a basilar artery (BA) aneurysm or a superior cerebellar artery (SCA) aneurysm. She was treated by endovascular therapy using detachable coils through the BA. The BA aneurysm was completely occluded, but part of a coil migrated into the BA from the SCA aneurysm during the procedure. The migrated coil was retrieved using a snare type endovascular retrieving device. The snare loop required concentric closure around the coil by simultaneous pulling of the corewire and a slight forward movement of the tip of the snare catheter. The SCA aneurysm was successfully occluded 1 week later using a shorter coil. © 1995, The Japan Neurosurgical Society. All rights reserved.