Abstract
Hybrid photon counting detectors have significantly advanced synchrotron research. In particular, the introduction of large cadmium telluride-based detectors in 2015 enabled a whole new range of high-energy X-ray measurements. This article describes the specifications of the new PILATUS4 cadmium telluride detector and presents results from prototype testing for high-energy powder X-ray diffraction studies conducted at two synchrotrons. The experiments concern time-resolved in situ solid-state reactions at MAX IV (Sweden) and fast-scanning X-ray diffraction computed tomography of a battery cell at the ESRF (France). The detector's high quantum efficiency up to 100 keV, combined with a maximum frame rate of 4000 Hz, enables fast data collection. This study demonstrates how these capabilities contribute to improved time and spatial resolution in high-energy powder X-ray diffraction studies, facilitating advancements in materials, chemical and energy research.
Author supplied keywords
Cite
CITATION STYLE
Donath, T., Trampari, S., Wagner, L., Jørgensen, M. R. V., Gjørup, F. H., Checchia, S., … Vaughan, G. (2025). Enhancing high-energy powder X-ray diffraction applications using a PILATUS4 CdTe detector. Journal of Synchrotron Radiation, 32(Pt 2), 378–384. https://doi.org/10.1107/S1600577525000566