Comparison of biometric measurements obtained by the Verion Image-Guided System versus the auto-refracto-keratometer

5Citations
Citations of this article
25Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Purpose: To compare the biometric measurements obtained from the Verion Image-Guided System to those obtained by auto-refracto-keratometer in normal eyes. Methods: This is a prospective, observational, comparative study conducted at the Asociación para Evitar la Ceguera en México I.A.P., Mexico. Three sets of keratometry measurements were obtained using the image-guided system to assess the coefficient of variation, the within-subject standard deviation and intraclass correlation coefficient (ICC). A paired Student t test was used to assess statistical significance between the Verion and the auto-refracto-keratometer. A Pearson’s correlation coefficient (r) was obtained for all measurements, and the level of agreement was verified using Bland–Altman plots. Results: The right eyes of 73 patients were evaluated by each platform. The Verion coefficient of variation was 0.3% for the flat and steep keratometry, with the ICC being greater than 0.9 for all parameters measured. Paired t test showed statistically significant differences between groups (P = 0.0001). A good correlation was evidenced for keratometry values between platforms (r = 0.903, P = 0.0001 for K1, and r = 0.890, P = 0.0001). Bland–Altman plots showed a wide data spread for all variables. Conclusion: The image-guided system provided highly repeatable corneal power and keratometry measurements. However, significant differences were evidenced between the two platforms, and although values were highly correlated, they showed a wide data spread for all analysed variables; therefore, their interchangeable use for biometry assessment is not advisable.

Cite

CITATION STYLE

APA

Velasco-Barona, C., Cervantes-Coste, G., Mendoza-Schuster, E., Corredor-Ortega, C., Casillas-Chavarín, N. L., Silva-Moreno, A., … Gonzalez-Salinas, R. (2018). Comparison of biometric measurements obtained by the Verion Image-Guided System versus the auto-refracto-keratometer. International Ophthalmology, 38(3), 951–957. https://doi.org/10.1007/s10792-017-0541-3

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free