Vortex-assisted liquid-liquid microextraction and indirect spectrophotometric determination of chromium(vi)

11Citations
Citations of this article
18Readers
Mendeley users who have this article in their library.

Abstract

A novel, simple, sensitive and selective method for the indirect spectrophotometric determination of chromium(vi) was developed on the basis of vortex-assisted liquid-liquid microextraction (VALLME) of an ion association pair formed between the triiodide anion (I3−) and cationic dye Astrazone Brilliant Red 4G (ABR). The indirect spectrophotometric determination of chromium(vi) was based on the redox-reaction of Cr(vi) with iodide and formed I3−. The triiodide anion was determined spectrophotometrically at 530 nm following VALLME of an ion association pair with ABR. The calibration plot was linear in the range 0.050-4.2 μg L−1 (R2 = 0.998), and limit of detection (LOD) was 0.015 μg L−1 for Cr(vi). The suggested procedure was successfully applied for the determination Cr(vi) in water samples (RSD 3.3-5.7%, recovery 97.7-102.8% when determining ultratrace levels Cr(vi) at 0.55-5.77 μg L−1), the accuracy of the method was assessed through a comparison with reference methods and certified reference materials.

Cite

CITATION STYLE

APA

Bazel, Y., & Riabukhina, T. (2018). Vortex-assisted liquid-liquid microextraction and indirect spectrophotometric determination of chromium(vi). RSC Advances, 8(62), 35360–35366. https://doi.org/10.1039/C8RA07514J

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free