Abstract
The incidence of kidney disease is increasing worldwide. Rapid and cost-effective approaches for early detection help prevent this disease. Neutrophil gelatinase-associated lipocalin protein (NGAL) is a novel biomarker for acute kidney injury (AKI) and chronic kidney disease (CKD). We aimed to develop a lateral flow strip (LFS) based on a lateral flow immunoassay method (LFIA), using latex microspheres (LMs) as a color labeling to detect NGAL in urine. The performance and potential of the developed LMs-LFS at a point-of-care (POC) testing were evaluated. The results showed that LMs-LFS successfully detected urinary NGAL within 15 min with high specificity without cross-reactivity to or interference from other endogenous substances in urine. The visual limit of detection (vLOD) was 18.75 ng/mL, and the limit of detection (LOD) was 1.65 ng/mL under the optimum condition. The LMs-LFS developed in this study showed a high correlation with the enzyme-linked immunosorbent assay (ELISA) method (R2 = 0.973, n = 60 urine specimens) for detecting NGAL in urine. The LMs-LFS remained stable for at least six months at room temperature. The LMs-LFS can be a rapid, sensitive, and specific tool for the diagnosis and follow-up of renal disorders at the POC.
Cite
CITATION STYLE
Tunakhun, P., Ngernpimai, S., Tippayawat, P., Choowongkomon, K., Anutrakulchai, S., Charoensri, N., … Daduang, J. (2024). A Highly Sensitive Lateral-Flow Strip Using Latex Microspheres to Detect NGAL in Urine Samples. ACS Omega. https://doi.org/10.1021/acsomega.4c04322
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.