Instability of total hip prostheses at rotational stress: A roentgen stereophotogrammetric study

63Citations
Citations of this article
16Readers
Mendeley users who have this article in their library.

Abstract

Roentgen stereophotogrammetric analysis (RSA) has been carried out on 24 total hip prostheses (24 patients), which were all painful at weight-bearing. the examinations were made with the hip at distraction, compression, external and internal rotation. One acetabular component showed instability only at distraction-compression and one showed instability only at rotation. Four femoral components showed instability at distraction-compression and at rotation, six only at rotation, but none showed instability only at distraction-compression. Thus rotational provocation demonstrated instability in all unstable femoral components; rotational stress must be of importance in mechanical loosening of the femoral component. © 1984 Informa UK Ltd All rights reserved: reproduction in whole or part not permitted.

Cite

CITATION STYLE

APA

Mjöberg, B., Hansson, L. I., & Selvik, G. (1984). Instability of total hip prostheses at rotational stress: A roentgen stereophotogrammetric study. Acta Orthopaedica, 55(5), 504–506. https://doi.org/10.3109/17453678408992946