Image quality improvement for medium and large field of view Elekta XVI scans

4Citations
Citations of this article
22Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Cone-beam computed tomography (CBCT) has become the standard imaging technique for many sites treated with Radiotherapy. The Elekta X-ray Volumetric Imaging (XVI) system allows for the acquisition of CBCTs with three different diameters; small (27 cm), medium (41 cm) and large (50 cm) (Elekta in XVI Corrective Maintenance Manual R5.0, Elekta, Stockholm, 2013). Images are used to ensure accurate patient positioning as well as checking for changes in the patient contour or internal geometry, and image quality must be high enough to achieve these goals. This paper describes two simple adjustments which can lead to improved image quality for medium and large field of view (MFOV and LFOV) scans. The first involves measuring a default distance in the XVI software (kV Source to Detector Distance) to improve spatial resolution and geometric scaling of both MFOV and LFOV scans. The second involves correcting a uniformity issue seen at our centre with LFOV scans, allowing us to use this FOV for assessing patient contour changes on large patients. The implementation and effects of both adjustments are explored in this paper, and workflows are proposed for optimising both parameters.

Author supplied keywords

Cite

CITATION STYLE

APA

Little, D. P. (2019). Image quality improvement for medium and large field of view Elekta XVI scans. Australasian Physical and Engineering Sciences in Medicine, 42(4), 1153–1164. https://doi.org/10.1007/s13246-019-00817-7

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free