Accurate transfer function determination for superconducting gravimeters

57Citations
Citations of this article
8Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The transfer function for the cryogenic gravimeter GWR-C021 operating at Membach (Belgium) has been experimentally determined by injecting known voltages into the control electronics of the system. The output of the gravimeter to the injected sine waves and step functions has been observed. This give a precise knowledge of the transfer function of the gravimeter. It allows one to reach a precision of better than 0.01 second in the phase response of the instrument, in agreement with the Global Geodynamics Project (GGP) requirements.

Cite

CITATION STYLE

APA

Van Camp, M., Wenzel, H. G., Schott, P., Vauterin, P., & Francis, O. (2000). Accurate transfer function determination for superconducting gravimeters. Geophysical Research Letters, 27(1), 37–40. https://doi.org/10.1029/1999GL010495