Non-enzymatic amperometric glucose screen-printed sensors based on copper and copper oxide particles

8Citations
Citations of this article
19Readers
Mendeley users who have this article in their library.

Abstract

Non-enzymatic amperometric glucose sensors have gained much attention in the past decade because of the better chemical and thermal stability and biocompatibility compared to conventional sensors based on the use of biomolecules. This study focuses on a novel copper and copper oxide-based glucose sensor synthesized by an electrodeposition technique through a rigorous protocol which reports an excellent analytical performance due to its structure and its increased active area. In addition, the linear response range, detection limit and sensitivity were 0.5–5.0 mmol L−1, 0.002 mmol L−1, 904 µA mmol−1 L−1 cm−2, respectively. Results show a reliable electrode as it is chemically stable, exhibits rapid and excellent sensitivity, and it is not significantly affected by coexisting species present in the blood samples; furthermore, it reports a maximum relative standard deviation error (RSD) of 6%, and showed long operating life as the electrode was used for thousand measurements of 4.0 mmol L−1 glucose solution during three days.

Cite

CITATION STYLE

APA

Guati, C., Gomez-Coma, L., Fallanza, M., & Ortiz, I. (2021). Non-enzymatic amperometric glucose screen-printed sensors based on copper and copper oxide particles. Applied Sciences (Switzerland), 11(22). https://doi.org/10.3390/app112210830