Abstract
Background: Insulin analogues used in the treatment of insulin-dependent diabetes may result in hypoglycaemia; however, these preparations show variable cross-reactivity in many insulin assays. In order to investigate a suspected insulin overdose, it is therefore essential to characterize the cross-reactivity of these preparations in insulin assays. Methods: The cross-reactivity of 12 commercial insulin preparations was assessed on the Beckman DxI analyser, by comparison with the insulin calibration curve corresponding to the midpoint of the dose-response curve. This was achieved by manually plotting the relative light units generated by the analyser for each preparation. Results: All 12 insulin preparations examined showed significant cross-reactivity with the Beckman DxI insulin assay, with 11 of the insulin preparations demonstrating cross-reactivity of 83% or greater. The lowest cross-reactivity was observed with detemir with a value of 47.6%. Conclusion: When investigating hypoglycaemia it is important to be aware of the cross-reactivity of the recombinant insulin analogues in insulin assays. This study has established the cross-reactivity of 12 preparations in the Beckman DxI insulin assay.
Cite
CITATION STYLE
Glenn, C., & Armston, A. (2010). Cross-reactivity of 12 recombinant insulin preparations in the Beckman Unicel DxI 800 insulin assay. Annals of Clinical Biochemistry, 47(3), 264–266. https://doi.org/10.1258/acb.2010.010002