The design and performance characterization of a new light-weight and compact X-ray scintillation detector is presented. The detectors are intended for use on the new I11 powder diffraction beamline at the third-generation Diamond synchrotron facility where X-ray beams of high photon brightness are generated by insertion devices. The performance characteristics of these detection units were measured first using a radioactive source (efficiency of detection and background count rate) and then synchrotron X-rays (peak stability, light yield linearity and response consistency). Here, the results obtained from these tests are reported, and the suitability of the design for the Diamond powder beamline is demonstrated by presenting diffraction data obtained from a silicon powder standard using a prototype multicrystal analyser stage. © International Union of Crystallography 2008.
CITATION STYLE
Tartoni, N., Thompson, S. P., Tang, C. C., Willis, B. L., Derbyshire, G. E., Wright, A. G., … Bell, A. M. T. (2007). High-performance X-ray detectors for the new powder diffraction beamline I11 at Diamond. Journal of Synchrotron Radiation, 15(1), 43–49. https://doi.org/10.1107/S0909049507046250
Mendeley helps you to discover research relevant for your work.