Clinical management of tumour volume changes in VMAT head & neck radiation treatment

4Citations
Citations of this article
11Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The impact of changing anatomy due to tumour shrinkage was assessed for a VMAT head & neck plan optimized according to our centre's planning protocol. A custom-built wax phantom accommodating ion chamber, MOSFET, EBT3 film and Fricke-xylenol orange-gelatin (FXG) gel dosimeters and a variable size bolus 'tumour' was used in the investigation. Results indicate that the practice of initiating a patient re-scan and verification plan is appropriate when a change in external body contour greater than 1 cm compared to the original anatomy is observed.

Cite

CITATION STYLE

APA

Alexander, K. M., Gooding, J., Schreiner, L. J., & Olding, T. (2017). Clinical management of tumour volume changes in VMAT head & neck radiation treatment. In Journal of Physics: Conference Series (Vol. 847). Institute of Physics Publishing. https://doi.org/10.1088/1742-6596/847/1/012038