Studying of Common Factors Affecting Reading of Lithium Fluoride Thermo-Luminescence Dosimeter Crystal

  • Omer M
  • Alsayyari A
  • Tamboul J
  • et al.
N/ACitations
Citations of this article
7Readers
Mendeley users who have this article in their library.

Abstract

The aim of this study was to study the technical factors affecting the reading of TLD (LiF:Mg:Ti) for radiation dosimetry. The method adapted was an exposure of LiF:Mg:Ti crystal with γ-ray with the following filters (Hp 0.07 and Hp 10), before and after usage of filter Hp (0.07), annealing, cooling by nitrogen flux and crystal morphology by scanning electron microscope. The collected data analyzed by EXCELL software showed that: TLD average count/second (C/s) with filter Hp (10) was greater than that obtained with filter Hp (0.07) with an amount of 1191 c/s (19%). Also the analysis showed that: the average C/s without filter Hp (0.07) was greater than that with filter by an amount of 994.5 C/s (16.4%), and there was proportional linear relationship; as the applied dose increased the TLD C/s increased significantly (R2 = 0.9) based on the equations: y = 885.1x − 286.5 and y = 860.5x + 856.8 for filtered and without filtered beam respectively...

Cite

CITATION STYLE

APA

Omer, M. A. A., Alsayyari, A. A., Tamboul, J. Y., Ali, R. B., Ahmed, A. A., & Abukonna, A. (2017). Studying of Common Factors Affecting Reading of Lithium Fluoride Thermo-Luminescence Dosimeter Crystal. Open Journal of Biophysics, 07(01), 14–24. https://doi.org/10.4236/ojbiphy.2017.71002

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free