Background: To understand the accuracy of volume calibration syringes used in China and compare the difference between new and old volume calibration syringes, technical testing was performed on volume calibration syringes in clinical lung function instruments. Materials and methods: A standard validator device (Model 1180, Hans Rudolph, USA) was used to perform leak testing and volume accuracy testing for calibration syringes. Sixteen volume calibration syringes from 8 brands (CareFusion in Germany, Vyaire in Germany, Yaeger in Germany, Vitalograph in the United Kingdom, MGC Diagnostics in the United States, U-Breath in Zhejiang, China, Wendi in Ningbo, Zhejiang, and Boya in Ningbo, China) were tested. Results: A total of 75% (12/16) of the volume calibration syringes passed the pressure decay leak test, 69% (11/16) of the volume calibration syringes passed the volume accuracy and repeatability test, and 56% (9/16) passed both tests; there was no significant difference in the total passing of the new and old volume calibration syringe quality tests (P > 0.05). Conclusions: A standard validator device should be used for both leakage tests and volume accuracy and repeatability tests to ensure the reliability of volume calibration syringes. It is suggested that the quality verification of volume calibration syringes should be regularly conducted to ensure the accuracy of the pulmonary function tests.
CITATION STYLE
Wu, Z., Peng, Y., Lin, K., Huang, R., Zheng, J., & Gao, Y. (2022). Quality inspection and result analysis of the spirometer calibration cylinder. BMC Pulmonary Medicine, 22(1). https://doi.org/10.1186/s12890-022-02010-1