Fracture of temporary femoral haemodialysis catheter: Our experience

2Citations
Citations of this article
14Readers
Mendeley users who have this article in their library.

Abstract

We report two similar cases of fracture and proximal migration of temporary femoral haemodialysis catheters. The two cases were encountered 6 years apart. These types of occurrences, especially in centres where catheter reuse is practiced, are not uncommon but seldom reported in the literature. Alert dialysis staff on both the occasions took remedial steps in time, which prevented embolisation of the catheters into the inferior vena cava. Both the catheters were removed successfully by the surgical team. Copyright 2013 BMJ Publishing Group. All rights reserved.

Cite

CITATION STYLE

APA

Cruz, S. D., Gupta, M., Kaur, R., & Gupta, S. (2013). Fracture of temporary femoral haemodialysis catheter: Our experience. BMJ Case Reports. https://doi.org/10.1136/bcr-2013-009625