In Overhauser enhanced magnetic resonance imaging (OMRI) for in vivo measurement of oxygen partial pressure (pO2), a paramagnetic contrast agent is introduced to enhance the proton signal through dynamic nuclear polarization. A uniform proton T1 is generally assumed for the entire region of interest for the computation of pO2 using OMRI. It is demonstrated here, by both phantom and in vivo (mice) imaging, that such an assumption may cause erroneous estimate of pO2. A direct estimate of pixel-wise T1 is hampered by the poor native MR intensities, owing to the very low Zeeman field (15-20 mT) in OMRI. To circumvent this problem, a simple method for the pixel-wise mapping of proton T1 using the OMRI scanner is described. A proton T1 image of a slice through the center of an SCC tumor in a mouse clearly shows a range of T1 distribution (0.2∼1.6 s). Computation of pO2 images using pixel-wise T 1 values promises oximetry with minimal artifacts by OMRI.
CITATION STYLE
Matsumoto, S., Utsumi, H., Aravalluvan, T., Matsumoto, K. I., Matsumoto, A., Devasahayam, N., … Krishna, M. C. (2005). Influence of proton T1 on oxymetry using overhauser enhanced magnetic resonance imaging. Magnetic Resonance in Medicine, 54(1), 213–217. https://doi.org/10.1002/mrm.20564