Abstract
In 2004, a combined system test was performed in the H8 beam line at the CERN SPS with a setup reproducing the geometry of sectors of the ATLAS Muon Spectrometer, formed by three stations of Monitored Drift Tubes (MDT). The full ATLAS analysis chain was used to obtain the results presented in this paper. The basic design performances of the Muon Spectrometer were verified. The stability of MDT calibration constants, the alignment system using optical devices and high energy tracks, as well as the intrinsic sagitta resolution of the Muon Spectrometer were studied and found to agree with expectations. The reconstruction of muon tracks using the combined information from both the Inner Detector and the Muon Spectrometer are also presented. © 2008 Elsevier B.V.
Author supplied keywords
Cite
CITATION STYLE
Adorisio, C., Aielli, G., Alexopoulos, T., Alviggi, M., Amelung, C., Anastopoulos, C., … Zimmerman, S. (2009). Study of the ATLAS MDT spectrometer using high energy CERN combined test beam data. Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 598(2), 400–415. https://doi.org/10.1016/j.nima.2008.09.031