Abstract
Minimized perfusion circuits (MPCs) have been criticized for insufficient air elimination. The deairing capabilities of a new MPC, including an ultrasound controlled deairing unit, were compared to a standard extracorporeal circuit (ECC) in a laboratory setup. During blood flow of 4.0l/min, we injected 30-cc air over a period of 30 s into the venous line of both systems (n = 10 measurements/15-min intervals). Air was detected during the first 2 min post injection using a dual-channel ultrasound bubble counter. Venous air bubble measurements were made after the MPC bubble trap and the ECC hard-shell reservoir, respectively. Arterial air bubble data were obtained after the arterial filters (40 μm). Venous bubble count was significantly (P < 0.01) reduced in the MPC group (5-250 μm, 681 ± 177; >40 μm, 288 ± 92) compared with the ECC group (5-250 μm, 19 272 ± 682; >40 μm, 7642 ± 520). After the arterial filter, minimal numbers of air bubbles (5-250 μm, 172 ± 59; >40 μm, 0) could be detected in the MPC group, but large amounts of air (5-250 μm, 16 194 ± 1072; >40 μm, 3732 ± 997) were measured in the ECC group. The air elimination of the modern MPC is superior to conventional ECC, which may result in a reduction of neurological complications. © 2007, Copyright the Authors.
Author supplied keywords
Cite
CITATION STYLE
Kutschka, I., Schönrock, U., Essawi, A. E., Pahari, D., Anssar, M., & Harringer, W. (2007). A new minimized perfusion circuit provides highly effective ultrasound controlled deairing. Artificial Organs, 31(3), 215–220. https://doi.org/10.1111/j.1525-1594.2007.00367.x