A new minimized perfusion circuit provides highly effective ultrasound controlled deairing

23Citations
Citations of this article
15Readers
Mendeley users who have this article in their library.

Abstract

Minimized perfusion circuits (MPCs) have been criticized for insufficient air elimination. The deairing capabilities of a new MPC, including an ultrasound controlled deairing unit, were compared to a standard extracorporeal circuit (ECC) in a laboratory setup. During blood flow of 4.0l/min, we injected 30-cc air over a period of 30 s into the venous line of both systems (n = 10 measurements/15-min intervals). Air was detected during the first 2 min post injection using a dual-channel ultrasound bubble counter. Venous air bubble measurements were made after the MPC bubble trap and the ECC hard-shell reservoir, respectively. Arterial air bubble data were obtained after the arterial filters (40 μm). Venous bubble count was significantly (P < 0.01) reduced in the MPC group (5-250 μm, 681 ± 177; >40 μm, 288 ± 92) compared with the ECC group (5-250 μm, 19 272 ± 682; >40 μm, 7642 ± 520). After the arterial filter, minimal numbers of air bubbles (5-250 μm, 172 ± 59; >40 μm, 0) could be detected in the MPC group, but large amounts of air (5-250 μm, 16 194 ± 1072; >40 μm, 3732 ± 997) were measured in the ECC group. The air elimination of the modern MPC is superior to conventional ECC, which may result in a reduction of neurological complications. © 2007, Copyright the Authors.

Cite

CITATION STYLE

APA

Kutschka, I., Schönrock, U., Essawi, A. E., Pahari, D., Anssar, M., & Harringer, W. (2007). A new minimized perfusion circuit provides highly effective ultrasound controlled deairing. Artificial Organs, 31(3), 215–220. https://doi.org/10.1111/j.1525-1594.2007.00367.x

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free