Variable-flip-angle 3D spiral-in-out turbo spin-echo imaging using concomitant gradient compensation and echo reordering at 0.55 T

2Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Purpose: To develop single-slab 3D spiral turbo spin echo (spiral SPACE) for 1-mm3 isotropic whole-brain T2-weighted imaging on a high-performance 0.55T scanner, with high scan efficiency from interleaved spiral-in-out trajectories, variable-flip-angle refocusing radiofrequency (RF) pulses, echo reordering, and concomitant-field compensation. Methods: A stack-of-spirals (in-out waveforms) turbo-spin-echo acquisition was implemented with T2-weighed contrast. Gradient infidelity was corrected using the gradient impulse response function (GIRF), and concomitant-field compensation was used to correct for phase errors among echoes and during the readout windows. To maintain a long echo train (˜600 ms) within each shot, variable-flip-angle refocusing RF pulses were generated using extended-phase-graph analysis. An echo-reordering scheme provided a smooth signal variation along the echo direction in k-space. Images from spiral SPACE with and without concomitant-field compensation were compared with those from Cartesian SPACE in phantoms and 6 healthy volunteers. Results: Phantom results demonstrated the improved performance of concomitant-field correction via sequence-based modifications and of GIRF–based trajectory estimation. Volunteer data showed that with concomitant-field correction and echo reordering, system imperfection associated image artifacts and blurring were substantially mitigated in spiral SPACE. Compared with Cartesian SPACE, spiral SPACE had an overall 15%–25% signal-to-noise ratio (SNR) improvement in both white matter and gray matter. Conclusion: A 3D spiral-in-out SPACE acquisition with variable-flip-angles, concomitant-field compensation, and echo-reordering was demonstrated at 0.55 T, showing promising gains in SNR, compared with Cartesian SPACE.

Cite

CITATION STYLE

APA

Wang, Z., Ramasawmy, R., Javed, A., Mugler, J. P., Meyer, C. H., & Campbell-Washburn, A. E. (2025). Variable-flip-angle 3D spiral-in-out turbo spin-echo imaging using concomitant gradient compensation and echo reordering at 0.55 T. Magnetic Resonance in Medicine, 93(4), 1741–1750. https://doi.org/10.1002/mrm.30380

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free