Abstract
The bioassay, frst order derivative UV spectrophotometry and chromatographic methods for assaying fuconazole capsules were compared. They have shown great advantages over the earlier published methods. Using the frst order derivative, the UV spectrophotometry method does not suffer interference of excipients. Validation parameters such as linearity, precision, accuracy, limit of detection and limit of quantitation were determined. All methods were linear and reliable within acceptable limits for antibiotic pharmaceutical preparations being accurate, precise and reproducible. The application of each method as a routine analysis should be investigated considering cost, simplicity, equipment, solvents, speed, and application to large or small workloads.
Author supplied keywords
Cite
CITATION STYLE
Corrêa, J. C. R., Reichman, C., Salgado, H. R. N., & Vianna-Soares, C. D. (2012). Performance characteristics of high performance liquid chromatography, first order derivative UV spectrophotometry and bioassay for fluconazole determination in capsules. Quimica Nova, 35(3), 530–534. https://doi.org/10.1590/S0100-40422012000300017