Heat losses from babies in incubators

47Citations
Citations of this article
27Readers
Mendeley users who have this article in their library.

Abstract

Simple methods by which the thermal environment provided by an incubator can be measured and analysed into conductive, convective, radiant, and evaporative components have been developed and applied to two commercial incubators working on different principles. Radiant heat losses were found to be considerable unless room temperature was unusually high. It is predicted that a naked newborn premature infant placed in such an incubator with the air at 34° C. (93° F.) will need to increase its oxygen consumption 25% above the basal level to maintain its body temperature at 37° C. if the room temperature is 20° C. (68° F.). The clinical significance of these findings is discussed and various methods of reducing or compensating for these losses are suggested.

Cite

CITATION STYLE

APA

Hey, E. N., & Mount, L. E. (1967). Heat losses from babies in incubators. Archives of Disease in Childhood, 42(221), 75–84. https://doi.org/10.1136/adc.42.221.75

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free