Left atrial appendage closure with the Amplatzer™ Cardiac Plug: Rationale for a higher degree of device oversizing at implantation

8Citations
Citations of this article
13Readers
Mendeley users who have this article in their library.

Abstract

Background: In left atrial appendage (LAA) closure, the correct sizing of the implantable devices is crucial. Data on the time-dependent changes in the shape and positioning of LAA occlusion devices are missing. We analyzed the results of 33 consecutive patients after implantation of an Amplatzer™ Cardiac Plug (ACP) LAA closure device to get more information on the optimal device sizing during implantation. Methods and results: Thirty-three consecutive patients were enrolled in this observational study. ACP implantation was guided by fluoroscopy and three dimensional transesophageal echocardiography (3-D TEE). Device sizing was based on the largest measured diameter of the intended landing zone adding 2–4 mm of device oversizing. Fluoroscopies were performed at 1 day after, and after 3 months, control 3-D TEE was performed 3 months after implantation. The stability of device positioning and shape was matched with the results of 3-D TEE. Patients’ mean age was 70.2 ± 8 years; mean CHA 2DS2VASc score was 3.8 ± 1.1. According to the manufacture’s classification, the post-implant degree of compression of the device-lobe was classified in three categories 1) undercompression “square-like shape” (1 patient); 2) optimal compression “tire-like shape” (20 patients), 3) overcompression “strawberry-like shape” (12 patients). Changes in the degree of device compression by more than one classification class occurred in 18/33 of our patients. A complete loss of device compression (“square-like shape”) was observed in 9 patients. Despite the changes in device compression, a complete closure of the LAA was achieved in 32/33 patients. Conclusions: There is a temporal change in shape and positioning of the ACP within 3 months after implantation. A late decompression of the ACP lobe was observed in 61% of our patients, leading to a complete loss in device compression in 27%. This observation may be the rationale for a higher degree of ACP oversizing during implantation.

Cite

CITATION STYLE

APA

Neuzner, J., Dietze, T., Paliege, R., Möller, M., Saeed, G., & Gradaus, R. (2015). Left atrial appendage closure with the AmplatzerTM Cardiac Plug: Rationale for a higher degree of device oversizing at implantation. Cardiology Journal, 22(2), 201–205. https://doi.org/10.5603/CJ.a2014.0063

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free