Use of stabilometric platform and evaluation of methods for further measurements--a pilot study.

3Citations
Citations of this article
24Readers
Mendeley users who have this article in their library.

Abstract

Posture in a still stance has been quantified by changes in the center of pressure (COP), in both anterior-posterior (A/P) and medial-lateral (M/L) directions and measured on a single force platform (Bertec PRO VEC 5.0). The purpose of this study was to estimate the variance in error and the intrasession test-retest reliability, and to determine which measures shall be taken for further measurements, especially with adults age 65 and older. We used two types of approximation for the reliability coefficient. Firstly, we used the equation according to Blahus (2) and secondly we used the Pearson's correlation coefficient for test-retest measurements. The findings allow us to say, among other things, that the tests of quiet standing Double Narrow Stance Eyes Open (DNSEO) and Double Narrow Stance Eyes Closed (DNSEC) are parallel, in the sense of parallel testing.

Cite

CITATION STYLE

APA

Jancová, J., & Tosnerová, V. (2007). Use of stabilometric platform and evaluation of methods for further measurements--a pilot study. Acta Medica (Hradec Králové) / Universitas Carolina, Facultas Medica Hradec Králové, 50(2), 139–143. https://doi.org/10.14712/18059694.2017.71