Less than 20% of patients with sensorineural hearing loss (HL) provided with the Vibrant Soundbridge (VSB) experience a progressive HL and warrant cochlear implantation (CI). The aim of this study was to identify possible predictors of progressive HL prior to VSB implantation. This retrospective study included all consecutive ears with sensorineural HL provided with the VSB between 1998 and 2016. The patient cohort was divided in a study group comprising patients who underwent CI (CI group) after years of VSB usage and those who did not require VSB replacement during the observational time (control group). Pre-and postoperative pure-tone audiometry thresholds were compared among the two groups. Fifteen out of 81 VSB devices (18.5%) required a CI. The CI group had higher preoperative air-conduction (AC) thresholds than the control group (64.3 ± 8.9 dB vs. 56.3 ± 12.9 dB; p = 0.007) at the time of the VSB implantation. On average, the CI group was significantly younger (39.1 ± 12.3 years vs. 52.6 ± 16.2 years; p = 0.003). In conclusion, VSB users with higher preoperative AC thresholds and younger age at the time of VSB implantation might be at risk for progressive HL within the upcoming eight years and need a further CI surgery. Preoperative counseling is particularly advisable in this patient group.
CITATION STYLE
Brkic, F. F., Baumgartner, W. D., Riss, D., Thurner, T., Liu, D. T., Gstöttner, W., & Vyskocil, E. (2022). Progressive Sensorineural Hearing Loss in Vibrant Soundbridge Users Requiring Cochlear Implantation. Journal of Personalized Medicine, 12(2). https://doi.org/10.3390/jpm12020191
Mendeley helps you to discover research relevant for your work.