A significant challenge in the production of Earth observation satellites is the precise alignment of the telescope optical components. We have developed a strategy to perform automated alignment of two-mirror telescopes for use in a realistic factory-based setting. A Ritchey–Chrétien telescope was used as an example. The secondary mirror was mounted on a high precision hexapod and its misalignment inferred from the Zernike coefficients for tilt, defocus, and coma, as measured by a phase-shifting interferometer. The required corrections to the position of the secondary mirror were implemented using an integral controller and alignment was achieved within minutes, compared to within days when using a manual alignment process. The Zernike coefficient for each aberration was reduced to within one standard deviation of the fluctuations due to residual instability (48 nm).
CITATION STYLE
Hampson, K. M., Gooding, D., Cole, R., & Booth, M. J. (2019). High precision automated alignment procedure for two-mirror telescopes. Applied Optics, 58(27), 7388. https://doi.org/10.1364/ao.58.007388
Mendeley helps you to discover research relevant for your work.