Abstract

Background: We compared the cell counting accuracy of the conventional method and the improved method by using Neubauer counting chamber. Methods: In the improved method, all the border cells were counted and then divided by two; while, in the conventional method, only border cells on the two boundaries (top and left) were counted. Results: About 55.814% of the samples showed more accurate results by improved counting method, about 38.372% had more accurate results by conventional counting method, and about 5.814% were counted with similar counting error by both methods. The improved method had significantly smaller counting error than conventional method (P

Cite

CITATION STYLE

APA

Zhang, M., Gu, L., Zheng, P., Chen, Z., Dou, X., Qin, Q., & Cai, X. (2020). Improvement of cell counting method for Neubauer counting chamber. Journal of Clinical Laboratory Analysis, 34(1). https://doi.org/10.1002/jcla.23024