Abstract
Study Design: Cadaveric study. Objective: To evaluate accuracy, radiation exposure, and surgical time of a new robotic-assisted navigation (RAN) platform compared with freehand techniques in conventional open and percutaneous procedures. Methods: Ten board-certified surgeons inserted 16 pedicle screws at T10–L5 (n = 40 per technique) in 10 human cadaveric torsos. Pedicle screws were inserted with (1) conventional MIS technique (L2–L5, patient left pedicles), (2) MIS RAN (L2–L5, patient right pedicles), (3) conventional open technique (T10–L1, patient left pedicles), and (4) open RAN (T10–L1, patient right pedicles). Output included (1) operative time, (2) number of fluoroscopic images, and (3) screw accuracy. Results: In the MIS group, compared with the freehand technique, RAN allowed for use of larger screws (diameter: 6.6 ± 0.6 mm vs 6.3 ± 0.5 mm; length: 50.3 ± 4.1 mm vs 46.9 ± 3.5 mm), decreased the number of breaches >2 mm (0 vs 7), fewer fluoroscopic images (0 ± 0 vs 108.3 ± 30.9), and surgical procedure time per screw (3.6 ± 0.4 minutes vs 7.6 ± 2.0 minutes) (all P 2 mm (0 vs 13), fewer fluoroscopic images (0 ± 0 vs 24.1 ± 25.8) (all P
Author supplied keywords
Cite
CITATION STYLE
Vaccaro, A. R., Harris, J. A., Hussain, M. M., Wadhwa, R., Chang, V. W., Schroerlucke, S. R., … Bucklen, B. S. (2020). Assessment of Surgical Procedural Time, Pedicle Screw Accuracy, and Clinician Radiation Exposure of a Novel Robotic Navigation System Compared With Conventional Open and Percutaneous Freehand Techniques: A Cadaveric Investigation. Global Spine Journal, 10(7), 814–825. https://doi.org/10.1177/2192568219879083
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.