Urease testing of mycobacteria with BACTEC radiometric instrumentation

3Citations
Citations of this article
6Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

A total of 140 mycobacterial isolates from patients treated at Fitzsimons Army Medical Center or the National Jewish Hospital and Research Center and from animal specimens submitted to the National Veterinary Services Laboratory were tested by using a urease procedure modified for use with a BACTEC model 301. Mycobacterial suspensions were prepared by using Middlebrook 7H10 Tween broth. Of the 98 mycobacteria isolates which were urease positive utilizing standard methodology, all were positive using the radiometric procedure. Similarly, all 42 urease-negative isolates were also negative employing the new methodology. Although maximum radiometric readings were observed at 48 h, all positive strains were readily identified 24 h after inoculation without sacrificing either test sensitivity or specificity. Thus, urease testing of mycobacteria, using the modified BACTEC radiometric methodology, was as sensitive, as specific, and more rapid than conventional methods.

Cite

CITATION STYLE

APA

Damato, J. J., Collins, M. T., & McClatchy, J. K. (1982). Urease testing of mycobacteria with BACTEC radiometric instrumentation. Journal of Clinical Microbiology, 15(3), 478–480. https://doi.org/10.1128/jcm.15.3.478-480.1982

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free